A-14 624074/07
A Specifications
Vt (tidal volume)
Adult 100 to 2000 ml 500 ml 10 ml for < 1000 ml
50 ml for > 1000 ml
Pediatric 20 to 300 ml 100 ml 1 ml for < 100 ml
10 ml for > 100 ml
Infant/neonatal -- -- --
Vtarget (target tidal volume)
Adult 100 to 2000 ml 500 ml 10 ml for < 1000 ml
50 ml for
1000 ml
Pediatric 20 to 300 ml 100 ml 1 ml for < 100 ml
10 ml for
100 ml
Infant/neonatal 2 to 200 ml 20 ml 0.1 ml for < 10 ml
1 ml for
10 ml
1. The standard setting is the default for a new patient for the first application of that specific
control. If the control setting is later changed, the new setting overrides the standard
setting.
2. With 2 ml option
3. Depending on how the ventilator is configured, you can see different time-related control
settings. See Appendix J for details.
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution