203
Function specications
Trigger active edge selection function
Function Allows you to select the active edge for trigger input from the EXT I/O (rising
or falling).
Operation mode DOWN/UP
DOWN: falling, UP: raising
Default setting DOWN
EOM
――
output method setting
Function Allows you to congure the instrument to hold the signal in the LO state for
the set amount of time after EOM
――
changes to LO and then restore it to the
HI state.
Operation mode HOLD/PULSE
HOLD : Holds the signal in the HI state for the period dened by
(analog measurement time and calculation time, and delay time).
PULSE: Holds the signal in the LO state for the set amount of time
after EOM
――
changes to LO and then restores it to the HI state.
Valid setting range 0.0001 s to 0.9999 s
Default setting HOLD, 0.0050 s
Delay setting from judgment result output to EOM
――
output
Function Allows you to set the delay time for EOM
――
(LO) output from judgment result
output.
Valid setting range 0.0000 s to 0.9999 s
Default setting 0.0000 s
Resetting judgment results
Function Allows you to set whether to reset the judgment results after EOM
――
changes
to HI.
Operation mode OFF/ON
OFF: Holds the judgment results until the next judgment;
ON : Resets the judgment results once EOM
――
changes to HI.
Default setting ON
System settings
Interface settings
RS-232C
Communications speed 9600 bps / 19200 bps / 38400 bps / 57600 bps (Default setting: 9600 bps)
Flow control OFF/ Hardware/ Software (Default setting: OFF)
Terminator CR+LF, CR (Default setting: CR+LF)
GP-IB
Address 00 to 30 (Default setting: 01)
Terminator LF, CR+LF (Default setting: LF)
10