Appendix B. Processing Flags
Code Description Outcome
sl Invalid kinetic profile.
The slope of the kinetic
curve is out of range for
the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
sp Invalid kinetic profile. A
spike was detected in
the kinetic profile.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
t Invalid fluorescent curve
profile. Assay
processing error.
Sample is invalid.
Sample must be retested.
Check all fluid connections. Contact Hologic
Technical Support.
ta The software has
reported an error in the
kinetic algorithm
calculation.
Sample is invalid.
Contact Hologic Technical Support.
ts The software has
reported that a kinetic
algorithm calculation is
invalid.
Sample is invalid.
Contact Hologic Technical Support.
ub The software has
reported a user bracket
control failure.
Control and associated samples are invalid.
Contact Hologic Technical Support.
x One or more of the
controls has been
invalidated, invalidating
all specimens currently
being processed in the
affected worklist.
Control and associated samples are invalid.
See additional results processing codes and
assay processing codes for more information.
Contact Hologic Technical Support.
z Invalid kinetic profile.
The total RLU value
was equal to zero for at
least one interval in the
kinetic curve.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Table 117. Results Processing Flags (Continued)