Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
287
Appendix B. Processing Flags
Code Description Outcome
n Invalid kinetic profile for
a specimen. The total
specimen RLU value is
less than the minimum
cutoff for the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
p The software has
reported an assay
processing error.
Sample is invalid.
See additional results processing codes and
assay processing codes for more information.
Contact Hologic Technical Support.
pn Invalid kinetic profile for
the negative calibrator.
The peak RLU value
has been reached either
too early or too late.
Calibrator replicate is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
pp Invalid kinetic profile for
the positive calibrator.
The peak RLU value
has been reached either
too early or too late.
Calibrator replicate is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
r The total RLU value is
greater than the
maximum cutoff value
for the assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
rm Invalid fluorescent curve
profile. Fluorescence
range falls outside
specified limits.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
rtf Invalid fluorescent curve
profile. Fluorescence
range falls outside
specified limits.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
rwb Invalid fluorescent curve
profile. Internal control
curve emergence falls
outside specified limits.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.