Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
286
Appendix B. Processing Flags
Code Description Outcome
fe Invalid or indeterminate
kinetic profile. The
glower analyte result is
positive but the flasher
analyte result is
equivocal.
Retest sample as an Equivocal according to
package insert instructions.
fge Invalid or indeterminate
kinetic profile. Both the
glower analyte and
flasher analyte have
results that are
equivocal.
Retest sample as an Equivocal according to
package insert instructions.
fl The software has
reported a results
interpretation error.
Sample is invalid.
Contact Hologic Technical Support.
g The calibrator glower
analyte curve is out of
range for the assay.
Calibrator replicate is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
ga The calibrator glower
RLU average is out of
range.
Calibrator replicate is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
ge Invalid or indeterminate
kinetic profile. The
flasher analyte result is
positive but the glower
analyte is equivocal.
Retest sample as an Equivocal according to
package insert instructions.
ic Invalid kinetic profile for
a specimen.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
m Invalid kinetic profile for
a specimen. The total
specimen RLU value
has reached the
maximum for the assay
too early.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Table 117. Results Processing Flags (Continued)