Appendix B. Processing Flags
Code Description Outcome
ch The control total RLU is
greater than the
maximum cutoff value.
Control and associated samples are invalid.
Check all fluid connections. Contact Hologic
Technical Support.
ci Invalid kinetic profile for
the control.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
cl The control RLU value
is less than the
minimum cutoff for the
assay.
Control and associated samples are invalid.
Check all fluid connections. Contact Hologic
Technical Support.
cv The software has
reported a results
interpretation error.
Sample is invalid.
Contact Hologic Technical Support.
e The software has
reported a results
processing error.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
f All samples in the
worklist are invalid, may
be due to multiple
calibrator errors or
multiple invalid kinetic
profiles.
Samples are invalid.
Check all fluid connections. Contact Hologic
Technical Support.
fa The average flasher
RLU value for a
calibrator is out of
range.
Calibrator replicate is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
fb Invalid kinetic profile.
The front-to-back ratio is
out of range for the
assay.
Sample is invalid.
Check all fluid connections. Contact Hologic
Technical Support.
Table 117. Results Processing Flags (Continued)