Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
354
Appendix C. System Messages
Code Message Possible Cause Solution
272.000.00008 LIS: System not
configured to
respond to query for
results.
Host query option is
not enabled.
Enable Host Query
Responding in the LIS
Configuration located
under the
Administration screen.
272.000.00009 Test Order for
Sample ID {0} lane
{1} position {2},
assay {3}, analytes
{4} was detected a
repeat. Test Order
was not created.
Repeat detection is
enabled. The
system attempted to
create the test order
and determined that
it was a repeat.
Manually override by
navigating to the
Sample Loading
screen and inspecting
the sample details.
272.000.00010 {0} in sample rack
{1}, position {2}
does not match
master lot of a
loaded assay
reagent kit.
System software
error.
1. Collect logs using
the Send Logs
feature (see Send
Logs on page
259).
2. Backup the
database.
3. Contact Hologic
Technical
Support.
272.000.00011 {0} in sample rack
{1}, position {2} is
not part of a
registered calibrator
master lot.
System software
error.
1. Collect logs using
the Send Logs
feature (see Send
Logs on page
259).
2. Backup the
database.
3. Contact Hologic
Technical
Support.
Table 118. System Messages (Continued)