Appendix C. System Messages
Code Message Possible Cause Solution
283.000.00012 Assay {0}:{1} Name
{2} version {3} not
loaded. Problems
detected while
loading the
associated DLL.
System software
error.
Contact Hologic
Technical Support.
283.000.00013 Assay {0}:{1} Name
{2} version {3}
cannot be used.
Problems detected
when the system
attempted to load
the Assay.
System software
error.
Contact Hologic
Technical Support.
356.000.00002 The assay
assembly returned
an invalid value.
System software or
database error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
356.000.00003 Too many
replicates of the
same level
detected in a set.
System software or
database error.
1. If the error is
persistent, shut
down and restart
when in the Setup,
Ready or Error
states.
2. Contact Hologic
Technical Support.
356.000.00004 An assay assembly
failed to load.
System software or
database error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
356.000.00006 Error when
mapping flags.
Please contact
technical support.
System software
error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
406.001.00001 Software exception
occurred
System software
error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.