Appendix C. System Messages
Code Message Possible Cause Solution
283.000.00002 Invalidated Sample
ID {1} for {2} assay
in rack {3} due to {0}
flag. Sample will
need to be retested.
See accompanying
message(s) to
determine cause.
All in-process
samples will be
invalidated.
1. Correct any flags.
2. Re-load and retest
all affected
samples.
3. If the problem
persists, contact
Hologic Technical
Support.
283.000.00003 {0} flag assigned to
Sample ID {1} for
{2} assay in lane {3}
position {4}.
See accompanying
message(s) to
determine cause.
1. Correct any flags
and continue.
2. If the problem
persists, contact
Hologic Technical
Support.
283.000.00004 {0} flag assigned to
Calibrator ID {1} for
{2} assay in lane {3}
position {4}.
See accompanying
message(s) to
determine cause.
1. Correct any flags
and continue.
2. If the problem
persists, contact
Hologic Technical
Support.
283.000.00005 {0} flag assigned to
Control ID {1} for {2}
assay in lane {3}
position {4}.
See accompanying
message(s) to
determine cause.
1. Correct any flags
and continue.
2. If the problem
persists, contact
Hologic Technical
Support.
283.000.00006 {0} flag assigned to
Sample ID {1} for
{2} assay in rack
{3}.
See accompanying
message(s) to
determine cause.
1. Correct any flags
and continue.
2. If the problem
persists, contact
Hologic Technical
Support.
283.000.00007 {0} flag assigned to
Calibrator ID {1} for
{2} assay in rack
{3}.
See accompanying
message(s) to
determine cause.
1. Correct any flags
and continue.
2. If the problem
persists, contact
Hologic Technical
Support.
Table 118. System Messages (Continued)