Appendix C. System Messages
Code Message Possible Cause Solution
272.000.00029 LIS transmission
terminated by user.
User cancelled LIS
during transmission.
1. If cancellation was
intentional, select
Save after altering
the LIS
configurations.
2. If cancellation was
not intentional,
enable LIS
connection (see
Enable LIS
Connection on
page 237).
277.000.00000 Scanning cannot
be started while the
door is open or
unlocked.
The issue could be
due to one of the
following:
l The TCR
Carousel
loading door is
open.
l Sensor is not
functioning
properly.
1. Firmly close the
TCR Carousel
loading door.
2. If the problem
persists, contact
Hologic Technical
Support.
282.000.00002 The {0} reagent has
expired.
The reagent
identified has
expired.
Load a new reagent to
continue assay
processing.
283.000.00001 Invalidated Sample
ID {1} for {2} assay
in lane {3} position
{4} due to {0} flag.
Sample will need to
be retested.
See accompanying
message(s) to
determine cause.
All in-process
samples will be
invalidated.
1. Correct any flags.
2. Re-load and retest
all affected
samples.
3. If the problem
persists, contact
Hologic Technical
Support.