Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
359
Appendix C. System Messages
Code Message Possible Cause Solution
272.000.00025 Reflex test {0} for
SID {1} was not
created as the
sample is invalid.
Invalidity reason:
{2}
An attempt was
made to create a
reflex test for a
sample which was
previously tested as
invalid.
1. See Appendix B.
Processing Flags
on page 272 for
explanation.
2. Take appropriate
steps to resolve
flag status.
3. Contact Hologic
Technical
Support.
272.000.00026 Test Order for
Sample ID {0} lane
{1} position {2},
assay {3} requires a
rack retainer.
Please add retainer
to rack and reload
the rack.
A sample rack was
loaded without the
sample rack
retainer.
Put retainer on the
sample rack and reload
the rack.
272.000.00027 Reflex test {0} for
SID {1} was not
created as the
sample is invalid.
Invalidity reason:
{2}
An attempt was
made to create a
reflex test for a
sample which was
previously tested as
invalid.
1. See Appendix B.
Processing Flags
on page 272 for
explanation.
2. Take appropriate
steps to resolve
flag status.
3. Contact Hologic
Technical
Support.
272.000.00028 Test Order for
Sample ID {0} lane
{1} position {2},
assay {3} requires a
rack retainer.
Please add retainer
to rack and reload
the rack.
A sample rack was
loaded without the
sample rack
retainer.
Put retainer on the
sample rack and reload
the rack.
Table 118. System Messages (Continued)