Appendix C. System Messages
Code Message Possible Cause Solution
272.000.00023 LIS order for
sample {0} rejected.
LIS test orders that
include dilution
factors are not
accepted by the
system.
An LIS test order
was received with a
dilution factor set for
the test order.
1. Consult the LIS
Host Guide and
ensure the
Dilution factor
sub-field is
correctly set to
one. Assistance
from your LIS
vendor may be
required.
2. If additional
assistance is
needed, contact
Hologic
Technical
Support.
272.000.00024 Error while
receiving data from
LIS.
An error occurred
while receiving data
from the LIS system.
1. Enable the ASTM
1381 and ASTM
1394 logging in
the LIS
configuration.
2. Retry attempt to
send test orders.
3. Contact LIS
vendor for
assistance
4. Contact Hologic
Technical
Support if
additional
assistance is
required.
Table 118. System Messages (Continued)