Code Message Possible Cause Solution
406.001.00010 Printer {0} is not
installed
No printer installed. No action necessary.
406.001.00011 Failed to set up
default printer {0}
Printer configuration
incomplete or
incorrect.
1. Ensure that the
printer is properly
connected and that
the Printer settings
on the System
Configuration
screen are correct.
2. If the problem
persists, contact
Hologic Technical
Support.
406.001.00012 Result ID: ({0})
failed its CRC
check and will not
be displayed.
Database integrity
check failure.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
406.001.00013 {0} results were not
exported due to
CRC check
failures. IDs: ({1})
Database integrity
check failure.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
406.001.00015 Failed to export
report. Reason: {0}
Message generated
as appropriate for
failure. Check
accompanying
messages if
available.
If the problem persists,
contact Hologic
Technical Support.