Appendix C. System Messages
Code Message Possible Cause Solution
406.001.00016 Assay Reagent Kit
ID:{0} (Assay: {1}) is
valid but does not
have a MasterLot
assigned to it.
Master Lot not
entered prior to
loading kit.
1. Unload the affected
Reagent Rack and
enter the Master Lot
information through
the Assay Reagent
Bay screen.
2. Reload the
Reagent Rack once
the Master Lot
information has
been properly
entered.
406.001.00017 The following result
IDs failed CRC
check and were not
displayed in the
report: {0}
Database integrity
check failure.
1. Shutdown and
restart.
2. If the problem
persists, contact
Hologic Technical
Support.
406.001.00018 Could not display
{0} results because
unknown assay '{1}'
was used to
process them
System software or
database error.
Contact Hologic
Technical Support.
406.001.00022 Maintenance task '
{0}' overdue
Required
maintenance task is
overdue. The
system is not able
to transition into the
Ready or Assay
Processing state
until the task has
been completed.
Complete Maintenance
task when in the Setup
state.