Appendix C. System Messages
Code Message Possible Cause Solution
406.001.00023 Maintenance task '
{0}' is due on {1}.
Required
maintenance task is
due in the time cited
in the message.
The system is not
able to transition
into the Ready or
Assay Processing
state until the task
has been
completed.
Complete Maintenance
task when in the Setup
or Ready state.
406.001.00025 Failed to merge old
configuration
settings.
System software or
database error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
406.001.00026 Value was out of
range for
configuration
setting {0} - {1}.
Please reconfigure
it.
Parameter setting is
out of range.
1. Reconfigure value
to specified
configuration
setting.
2. If the problem
persists, contact
Hologic Technical
Support.
406.001.00027 Configuration
setting file {0} is
invalid or corrupted
System software or
database error.
1. Shutdown and
restart.
2. Contact Hologic
Technical Support.
406.001.00028 Maintenance task '
{0}' did not auto-
start. Task is now
rescheduled to {1}.
Maintenance task
did not start at the
scheduled time.
Check associated
messages for
reason the task
failed to start.
1. If possible, correct
the cause of the
task failure listed in
the accompanying
message.
2. Retry maintenance
task.
3. If the problem
persists, contact
Hologic Technical
Support.