Panther System
®
Operator’s Manual
AW-17791-001 Rev. 002 (EN)
367
Appendix C. System Messages
Code Message Possible Cause Solution
406.001.00029 Sample Loading
defaults set: Assay
{0} Analytes {1}
N/A No action necessary.
406.001.00030 Sample Loading
defaults set:
Replicates {0}
N/A No action necessary.
406.001.00031 Required
maintenance is
suppressed until
{0}.
The required
maintenance task
has been
rescheduled. The
system is not able
to transition into the
Ready or Assay
Processing state
until the task has
been completed.
1. Complete
Maintenance task
when in the Setup
or Ready state.
2. If the problem
persists, contact
Hologic Technical
Support.
406.001.00032 Required
maintenance is
suppressed until
{0}.
The required
maintenance task
has been
rescheduled. The
system is not able
to transition into the
Ready or Assay
Processing state
until the task has
been completed.
1. Complete
Maintenance task
when in the Setup
or Ready state.
2. If the problem
persists, contact
Hologic Technical
Support.
406.001.00033 Required
maintenance is
now enabled.
N/A No action necessary.
406.001.00034 Required
maintenance is
now enabled.
An FSE disabled
the required
maintenance and it
is now re-enabled.
No action necessary.
Table 118. System Messages (Continued)