Appendix C. System Messages
Code Message Possible Cause Solution
407.010.00001 Failed to shut down
instrument.
An abnormal
termination of the
system was
detected.
1. Shut down and
restart the Panther
System Computer
using the button on
the computer.
2. If the problem
persists, contact
Hologic Technical
Support.
407.011.00001 Event log purge
configuration edited
Event log purge
configuration was
changed.
No action necessary.
407.011.00002 Maintenance log
purge configuration
edited
Maintenance log
purge configuration
was changed.
No action necessary.
407.011.00003 Activity log purge
configuration edited
Activity log purge
configuration was
changed.
No action necessary.
407.011.00004 Event logs purged Purge event logs
was selected.
No action necessary.