Configuration 
84  UDC3500 Universal Digital Controller Product Manual  3/07 
Function Prompt  
Lower Display 
Selections or  
Range of Setting  
Upper Display 
Parameter  
Definition 
 
LO SEL 
(See Note 2) 
INPUT LOW SELECT WITH RATIO AND BIAS—
This selection specifies the PV or SP as the lower of 
Input A or Input B. The following formula applies: 
 
Alg1 = lower of (Input A x Ratio A + Bias A) or (Input B x Ratio B + Bias B) 
 
√MuDIV 
(See Note 1) 
MULTIPLIER DIVIDER WITH SQUARE ROOT—
The following formula applies: 
Alg1 = K * Sq.Rt. {(Input A x Ratio A + Bias A) x (Input C x Ratio C + Bias C) / (Input B * Ratio B + Bias 
B)} 
 x (Calc Hi – Calc Lo) + Alg1Bias 
  See Figure 3-1 at the end of this section for an example of Mass Flow 
Compensation using the Multiplier/Divider Algorithm. 
 
√MULT 
(See Note 1) 
MULTIPLIER WITH SQUARE ROOT—The following 
formula applies: 
Alg1 = K x Sq.Rt. {(Input A x Ratio A + Bias A) x (Input B x Ratio B + Bias B) x  
(Input C x Ratio C + Bias C)} x (Calc Hi – Calc Lo) + Alg1Bias 
 
MuDIV  
(See Note 1) 
MULTIPLIER DIVIDER—The following formula 
applies: 
Alg1 = K x [{(Input A x Ratio A + Bias A) x (Input C x Ratio C + Bias C)} / (Input B x Ratio B + Bias B)] 
x (Calc Hi – Calc Lo) + Alg1Bias 
 
MULT  
(See Note 1) 
MULTIPLIER—The following formula applies: 
Alg1 = K x [(Input A x Ratio A + Bias A) x (Input C x Ratio C + Bias C) x (Input B x Ratio B + Bias B)] 
x (Calc Hi – Calc Lo) + Alg1Bias 
 
CARB A 
CARBON POTENTIAL A—Make this selection if 
you have a Cambridge or Marathon monitor type 
Zirconium Oxide sensor. It should also be used if 
using an Automotive probe (no thermocouple). This 
algorithm requires a temperature range within the 
region of 1500 to 2000°F. See 
Carbon/Oxygen/Dewpoint Notes. 
 
CARB B 
CARBON POTENTIAL B—Make this selection if 
you have a Corning type Zirconium Oxide sensor. 
This algorithm requires a temperature range within 
the region of 1500 to 1800°F. See 
Carbon/Oxygen/Dewpoint Notes. 
 
CARB C 
CARBON POTENTIAL C—Make this selection if 
you have an A.A.C.C. type Zirconium Oxide sensor. 
This algorithm requires a temperature range within 
the region of 1500 to 1900°F. See 
Carbon/Oxygen/Dewpoint Notes.