Troubleshooting/Service 
3/07  UDC3500 Universal Digital Controller Product Manual  327 
7.7.8 Procedure #8 – Analog Input 
Table 7-12 explains how to troubleshoot an Analog Input failure 
Table 7-12  Troubleshooting an Analog Input Failure 
Step  What to do  How to do it 
1 
Check Input Configuration.  Check if the input configurations are correctly 
set for the kind of sensor attached to the input 
terminals. See Section 
3.15 (Input 1) through 
Section 3.19 (Input 5). 
2 
Check input wiring and external 
resistor assemblies. 
See the Input Wiring Diagrams in Section 2.7 
and confirm that the instrument is properly 
connected to the sensor. 
Thermocouple, Milliamp, 0 to 10 Volt and –1 to 
1 Volt input types all require that external 
resistor assemblies be connected to the input 
terminals. These are provided with your 
instrument based upon the Model Number 
ordered. See the Input Wiring Diagrams in 
Section 
2.7 for installation information. 
3 
Check interconnection wiring 
If the actual sensor does not come directly to 
the controller but is instead connected via one 
or more intermediate junction panels, which in 
turn are connected to the controller, then 
check the continuity of the sensor to the 
controller. Check the tightness of the screws 
or connectors at the junction panels. 
4 
Check Input Signals.  Turn off power to the instrument (more than 
one switch may be necessary). Using a 
multimeter, measure the actual signal present 
at the rear terminals to ensure that it is within 
the allowed input range as shown in Section 
5.2. 
5 
Change the Input Type in order 
to restore Factory Calibration. 
See Section 5.6. 
6 
Replace Input Board. 
Installation instructions provided with new 
board. 
7 
Replace Controller.