14 Templates
917 Coulometer
■■■■■■■■
117
14.2.1 Editing result templates
List of result templates: Result template New / Edit
Result name
The result name is the text which will be shown in the results display and
in the report.
Entry max. 24 characters
Default value R
Calculation formula
Shows the calculation formula. A special editor is opened for the defini-
tion (see chapter 28.5.3, page 255).
Entry max. 100 characters
Default value empty
Decimal places
Number of decimal places used to display the result.
Input range 0 - 5
Default value 2
Result unit
The result unit is displayed and saved together with the result.
Entry max. 10 characters
Selection % | mol/L | mmol/L | g/L | mg/L | mg/mL |
mg/100 g | ppm | g | mg | µg | mL | µL | mg/
piece | °C | µg/min | mL/min | µL/min
Default value %
[Note]
Entering a note on the calculation.
[Note for wizard]
Enter a note on the calculation. This note is displayed when loading the
result template.
[Result options]
Defining additional settings for the calculation.
"Edit result template / Result options" dialog
In the dialog Edit result template / Result options, settings for how to
process the calculated result are defined.