21 Determination run
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
165
After the confirmation of the sample size, the live curve of the titra-
tion is displayed.
Sequence of the determination
ï¿ After a determination is started, a check is made as to
– whether all of the devices required are connected and ready for
operation.
– whether the sensor is present.
– whether the monitoring intervals for the sensor and common
variables are still valid.
– if a sample assignment table is activated (see chapter 14.1.2,
page 114):
whether the entered sample identification is defined in the sam-
ple assignment table. If this is the case, then the assigned
method will be loaded automatically and the actual determina-
tion will be started.
21.2 Performing a sample series
You can use the following functions with sample series:
ï¿ Statistics for multiple determinations
The statistics calculations are enabled in the method options (see chap-
ter 15.5, page 132).
ï¿ Sample table
You can enter the sample data for an entire sample series in the sample
table (see chapter 20, page 156).
Sequence of the sample series
ï¿ After a determination is started, a check is made as to
– whether all of the devices required are connected and ready for
operation.
–
– whether the sensor is present.
–
– whether the monitoring intervals for the sensor and common
variables are still valid.
– if a sample assignment table is activated (see chapter 14.1.2,
page 114):
whether the entered sample identification appears in the sample
assignment table. If this is the case, then the assigned method
will be loaded automatically and the actual determination will be
started. If a different method has been entered in the sample
table, then it will be ignored.