4.2 Setting up the titration cell for coulometry
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
917 Coulometer
4.2.3 Mounting the addition and aspiration tube
Figure 6 Mounting the addition and aspiration tube
Insert the addition and aspiration tube as follows into the titration cell:
1
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the 6.1437.000 stopper.
2
Insert the stopper into the 6.1439.010 addition and aspiration tube.
3
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
to the ground joint of the addition and aspiration tube.
4
Insert this assembly into the ground-joint opening.
5
Connect the tubing for the reagent addition at the upper connector
of the addition and aspiration tube (5).
6
Connect the tubing for the aspiration of the titration cell at the lower
connector of the addition and aspiration tube (6).
Details regarding how to connect the addition tubing and the aspiration
tubing can be found in the tutorial for the 917 Coulometer.