917 Coulometer
Parameter description
Stirring rate
Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 31.3, page 295.
The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:
ï¿ "+": counterclockwise rotation
ï¿ "–": clockwise rotation
Input range –15 - 15
Default value 8
NOTICE
If a stirrer is used in the determination run, it can still be manually con-
trolled at the same time. For example, you can use the manual control
to reduce the stirring rate of a stirrer which has been started in the
determination run.
27.5 Remote
Manual control â–¶ Remote
With the function [Remote], you can define manual output signals to the
remote interface of a connected Remote Box. It is not possible to define
input signals; the current status of the input lines is, however, displayed.