15.5 Method options
■■■■■■■■■■■■■■■■■■■■■■
138
■■■■■■■■
917 Coulometer
Monitoring sample size limits
on | off (Default value: off)
If this parameter is activated, then the sample size will be monitored.
Lower limit
Input range –999999999 - 9999999999
Default value 0
Upper limit
Input range –999999999 - 9999999999
Default value 9999999999
Display message
on | off (Default value: off)
If this parameter is activated, a message is displayed in case the limits are
infringed. You can select whether you want to continue with the determi-
nation or cancel the run.
If this parameter is deactivated, then the message will be entered only in
the message list of the determination.
Timeout
The message is displayed during this time. Afterwards, the determination
is continued automatically.
Input range 0 - 999999 s
Default value 30 s
Selection off
off
The determination will not be continued until after the message has
been confirmed.
15.5.4 Method properties
Main dialog: Edit parameters Method options Properties
Detailed information concerning the method is displayed in the Method
options / Properties dialog and the methods can be signed electroni-
cally here: