28 Parameter
917 Coulometer
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
229
■ "–": clockwise rotation
Input range –15 - 15
Default value 8
Switch off automatically
on | off (Default value: on)
If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.
28.2 Bromine index determination (BRC)
28.2.1 Control parameters
The control parameters for the endpoint are defined under [Control
parameters].
Endpoint at
Measured value for the endpoint.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Input range –1,250.0 - 1,250.0 mV
Default value 200.0 mV
Selection off
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection slow | optimal | fast | user
Default value optimal
slow
For samples with slow bromine addition.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For uncritical samples with fast bromine addition.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 10, page
230.

Table of Contents

Other manuals for Metrohm 917

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 and is the answer not in the manual?

Metrohm 917 Specifications

General IconGeneral
BrandMetrohm
Model917
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals