EasyManua.ls Logo

Metrohm 917 - Reports (REPORT)

Metrohm 917
338 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
■■■■■■■■■■■■■■■■■■■■■■
28 Parameter
917 Coulometer
■■■■■■■■
261
Variable Description
RMC Correlation coefficient for RM
Results, statistics
R# Result (# = 1 - 9)
SMN# Mean value of result R# (# = 1 - 9)
SSA# Absolute standard deviation of SMN# (# = 1 - 9)
SSR# Relative standard deviation of SMN# in % (# = 1 - 9)
SNR# Number of results from which the mean value SMN# was calculated (# = 1 - 9)
SSD Number of determinations that are carried out for the statistics calculations
SNT Statistics status (statistics activated = 1, statistics deactivated = 0)
Common variables
CV# Common variable (# = 1 - 25)
System variables
%RN Sample number, i.e. number that is increased by one at each start of a determination
%SC Start counter, see dialog More determination data / Properties
%AS Autostart status (autostart activated = 1, autostart deactivated = 0)
%AC Autostart counter, i.e. number of autostarts already carried out
%AD Autostart setpoint counter, i.e. number of autostarts to be carried out
%SS Status of the sample table (sample table activated = 1, sample table deactivated = 0)
%SL Number of the current sample table line
%SE Last sample in the sample table processed (yes = 1, no = 0)
28.6 Reports (REPORT)
The reports to be printed out are defined with the command REPORT. A
maximum of ten reports can be defined for each command. The following
reports can be inserted, among others:
Result report
Calculation report
Parameter report
System reports (system settings, sensor list, GLP data etc.)
Method reports (method run, method parameters etc.)