16 Electronic signatures
917 Coulometer
■■■■■■■■
143
16 Electronic signatures
16.1 Signing methods/determinations electronically
Methods: Main dialog Edit parameters Method
options Properties Sign
Determinations: Results dialog More data Properties Sign
You can sign methods or determinations in the dialog Electronic signa-
ture. The following information is saved along with each electronic signa-
ture:
User (complete name)
Date and time
Reason
Comment
Methods can only be signed if they have the status saved. The status of a
method can be found in the dialog Method options / Properties.
Signature
Selection of the level at which the determination should be signed.
Selection Review (signature level 1) | Release (signature
level 2)
A maximum of three signatures from up to three
different users can given at each level. However, it is
not possible for a user to sign at both level 1 and
level 2.
Review (signature level 1)
If the method has never been signed before, then only level 1 can be
selected.
Release (signature level 2)
If the method has been signed three times at level 1 or already once at
level 2 then only level 2 can be selected.
User
Identification of the user who signs.
Password
Password of the user.