28.1 Coulometric Karl Fischer titrations (KFC)
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
226
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
917 Coulometer
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Input range 1 - 999999 s
Selection off
Default value off
Delay reconditioning
on | off (Default value: off)
If this parameter is activated, the conditioning will not start before all the
method commands have been processed. Otherwise, the conditioning will
immediately start after the titration command.
NOTICE
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.
28.1.5 Cell
The parameters for the titration cell are defined under [Cell].
Generator electrode
Type of generator electrode.
Selection without diaphragm | with diaphragm
Default value without diaphragm
without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.
Generator current
Polarization current at the generator electrode.
Selection 100 mA | 200 mA | 400 mA | auto
400 mA
Default value, when Generator electrode = without diaphragm.