28 Parameter
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
225
Drift value
This parameter can only be edited with Drift correction = manual.
Drift for manual drift correction.
Input range 0.0 - 99.9 µg/min
Default value 0.0 µg/min
Measured value display
on | off (Default value: off)
If this parameter is activated, the currently measured value is displayed
during the conditioning.
Automatic start [Automatic start]
NOTICE
If you use the automatic start function, then we recommend that the
start drift be reduced and that a stabilizing time be defined.
Recommendation: start drift ≤ 10 µg/min / stabilizing time ≥ 30 s
Automatic start
on | off (Default value: off)
If this parameter is activated, the determination is automatically started
when the measured voltage suddenly changes. This setting will be ignored
as long as the working medium has not been conditioned.
Threshold value
If the change in voltage is higher than this value, the determination will
automatically be started.
Input range 0 - 999 mV
Default value 50 mV
Conditioning options [Cond. options]
Stabilizing time
Waiting time, during which the measured drift has to be smaller than the
start drift defined until Conditioning OK is displayed. The start drift is
defined in the dialog Edit command / Conditioning.
Input range 0 - 999999 s
Default value 0 s