EasyManuals Logo

Metrohm 917 User Manual

Metrohm 917
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #111 background imageLoading...
Page #111 background image
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
12 GLP manager
917 Coulometer
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
97
ï¿­ Select the hardware used for the test (see "Dialogs "GLP test
"Measurement" / Hardware" and "GLP test "Titration" / Hard-
ware"", page 99).
ï¿­ Tap on the fixed key [ ].
The dialog GLP tests HW/SW / GLP test "Measurement"
appears again.
4
Define the test interval
ï¿­ Tap on [GLP test interval].
ï¿­ Enter the time interval or the date for the next GLP test (see "Dia-
logs "GLP test "Measurement" / Test interval" and "GLP test
"Titration" / Test interval"", page 98).
ï¿­ Tap on the fixed key [ ].
12.3.1 Parameter description
Dialogs "GLP tests HW/SW / GLP test "Measurement"" and
"GLP tests HW/SW / GLP test "Titration""
Method
Method the GLP test has been carried out with.
Entry 32 characters maximum
Selection Selection of methods stored in the internal
memory
User
User who carried out the GLP test.
Entry 24 characters maximum
Selection Selection of configured users
Test date
Date on which the GLP was carried out.
Format: YYYY:MM:DD
Test result
Result of the test.
Selection Test OK | Test not OK
Default value Test OK
Comment
Entry 24 characters maximum

Table of Contents

Other manuals for Metrohm 917

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 and is the answer not in the manual?

Metrohm 917 Specifications

General IconGeneral
BrandMetrohm
Model917
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals