15.5 Method options
■■■■■■■■■■■■■■■■■■■■■■
140
■■■■■■■■
917 Coulometer
The following information is only shown if the method has been signed at
level 2:
■ Released by
User who signed the method at level 2.
■ Released on
Date and time at which the method has been signed at level 2.
■ Reason
Reason for the signing of the method.
■ Comment
Comment for the signing of the method.
[Delete signature]
Delete all of the signatures for the method or the determination. This but-
ton is inactive if level 2 has not yet been signed (see chapter 16.2, page
144).
[Sign]
Sign the method or determination. This button is active only when work-
ing with activated login function and password (see chapter 16.1, page
143).
15.5.5 Note
Main dialog: Edit parameters ▶ Method options ▶ Note
In this dialog, you can enter a short text, e.g. important information about
carrying out the determination.
[Display options]
Definition when the note is being displayed.
Automatically after loading the method
on | off (Default value: off)
If this parameter is activated, the note is displayed when loading the
method. It can otherwise only be read in this dialog.
15.5.6 Saving a determination automatically
Main dialog: Edit parameters ▶ Method options ▶ Save automat.
In the dialog Method options / Save automatically, you can specify
whether the determination is to be saved automatically and/or whether a
PC/LIMS report is to be created.
Save determination automatically
on | off (Default value: off)
If this parameter is activated, then the determination will be saved auto-
matically.