EasyManua.ls Logo

Metrohm 917 - Cell

Metrohm 917
338 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
■■■■■■■■■■■■■■■■■■■■■■
28 Parameter
917 Coulometer
■■■■■■■■
235
NOTICE
This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.
Example: Emptying the titration cell followed by adding new working
medium.
28.2.5 Cell
The parameters for the titration cell are defined under [Cell].
Generator electrode
Type of generator electrode.
Selection without diaphragm | with diaphragm
Default value with diaphragm
without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.
Generator current
Polarization current at the generator electrode.
Selection 100 mA | 200 mA | 400 mA | auto
Default value auto
auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint.
Reagent monitoring
Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System Reagents.
Selection Selection of configured reagents | off
Default value off
off
The reagent monitoring is not active.

Table of Contents

Other manuals for Metrohm 917

Related product manuals