ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
22 Live modifications
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
169
The live dialog is displayed once again.
22.3 Live display
Main dialog: Live display
The current method and the system status are displayed in the title line.
The currently running command is displayed directly underneath the title
line.
The following functions are available to you while a determination is run-
ning:
ï¿ [Hold]
Pause the determination. The label changes to [Continue]. This can be
used to continue the run.
ï¿ [Cancel command]
Cancel the current method command. The next command is started.
ï¿ [View] (only for titrations/measurements)
To modify the settings for the curve and measured value display, see
below.
ï¿ [Sample data] (visible only if a titration or a measurement is not cur-
rently running)
Switch to the main dialog in order to edit the sample data.
ï¿ [Edit parameters]
Edit the method options and live parameters (see chapter 22.4, page
170).
ï¿ [Results]
Display the results dialog (see chapter 23, page 172).
Dialog "View"
Main dialog: Live display â–¶ View
View
Type of curve display.
Selection Curve | Curve & meas. values | Measured val-
ues | Curve & sample data
Default value Curve & meas. values
Curve
Only the curve is displayed.
Curve & meas. values
The measured values are displayed to the right next to the curve.
Measured values
Only the measured values are displayed.
Curve & sample data
The sample data is displayed to the right next to the curve.