ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
15 Methods
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
133
able for the mean value is defined in the result options (see "Dialog "Edit
calculation / Result options"", page 252).
Number of determinations
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination has to be added to the determination
series, because one determination has been incorrect, for example, then
this can be accomplished in the statistical overview (see chapter 24, page
188).
Input range 2 - 20
Default value 3
[Save automat.]
Define the settings for automatically saving the determination (see chap-
ter 15.5.6, page 140).
[Properties]
Display the properties of the current method and sign electronically (see
chapter 15.5.4, page 138).
[Note]
Enter a note on the method (see chapter 15.5.5, page 140).
[Start/Stop options]
Edit functions which are carried out immediately after the method is
started or stopped, see chapter Start options, page133 and chapter Stop
options, page 134.
[Sample data]
Specify method-specific settings for the sample data, e.g. designation of
the sample identifications, limits for the sample size (see chapter 15.5.3,
page 135).
15.5.1 Start options
Main dialog: Edit parameters â–¶ Method options â–¶ Start/Stop
options â–¶ Start options
In the dialog Method options / Start options, you can configure the
functions which are carried out when a determination is started.
Method check at start
on | off (Default value: on)
If this option is activated, then the following tests are carried out at the
method start.
ï¿ Are the required control and peripheral devices connected?