EasyManuals Logo

Metrohm 917 User Manual

Metrohm 917
338 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #196 background imageLoading...
Page #196 background image
23.5 Loading a determination
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
182
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
917 Coulometer
Selection Result 1…9
The numbering of the results corresponds to the
order in which the results were calculated in the
determination run.
Sort by
Selection according to which criterion the list is to be sorted.
Selection Identification 1 | Identification 2 | Determ.
time | Result | Method | Sample size | User
Default value Determ. time
Identification 1
Sort alphabetically according to the first sample identification.
Identification 2
Sort alphabetically according to the second sample identification.
Determ. time
Sort chronologically according to the determination time.
Result
Sort according to the numerical value of the result.
Method
Sort alphabetically according to the method name.
Sample size
Sort according to the numerical value of the sample size.
User
Sort alphabetically according to the user who carried out the determi-
nation.
Sort sequence
Selection ascending | descending
Default value descending
[Filter]
Specify the filter criteria for the list.
Filter criteria
Determination list: Properties â–¶ Filter

Table of Contents

Other manuals for Metrohm 917

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 and is the answer not in the manual?

Metrohm 917 Specifications

General IconGeneral
BrandMetrohm
Model917
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals