27 Manual control
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
213
The volume dosed is displayed. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.
3
Fill the cylinder
ï¿ Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
NOTICE
If you exit the dialog with the fixed key [ ], the dosing cylinder is
filled automatically.
Parameter description
Dosing rate
Rate at which it is dosed.
Input range 0.01 - 166.00 mL/min
Selection maximum | dynamic
Default value maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 31, page 294).
dynamic
This setting can only be selected when the dosing cylinder is filled. The
dosing is being carried out faster and faster until the maximum dosing
rate is reached (starting with 1 mL/min and then doubling the dosing
rate every 1.5 s).
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 - 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
31, page 294).