28.1 Coulometric Karl Fischer titrations (KFC)
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
222
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
917 Coulometer
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see chapter 28.1.3,
page 223) always lead to a stop, even if the stop criterion has not been
reached.
Selection drift | time | rel. drift | off
Default value rel. drift
drift
The titration is stopped when the stop drift has been reached.
time
The titration is stopped if the endpoint has been exceeded during a
certain time period (Delay time).
rel. drift
The titration is stopped when the sum of the drift at the start of the
titration and the relative stop drift has been reached.
off
The titration will not be stopped until the stop conditions have been
fulfilled.
Stop drift
This parameter can only be edited with Stop criterion = drift.
The titration is stopped when the endpoint and the stop drift have been
reached.
Input range 1 - 999 µg/min
Default value 5 µg/min
Delay time
This parameter can only be edited with Stop criterion = time.
When the endpoint has been reached, the specified time is allowed to
elapse after the last dosing and the titration is then stopped.
Input range 0 - 999 s
Default value 10 s
Relative stop drift
This parameter can only be edited with Stop criterion = rel. drift.
The titration is stopped when the endpoint and the sum of the drift at the
start of the titration and the relative stop drift have been reached.

Table of Contents

Other manuals for Metrohm 917

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 and is the answer not in the manual?

Metrohm 917 Specifications

General IconGeneral
BrandMetrohm
Model917
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals