28 Parameter
917 Coulometer
ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­ï¿­
239
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range 0 - 999999 s
Default value 0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range 0 - 999999 s
The default value depends on the measuring mode.
Selection off
off
The measurement is continued endlessly.
Stop meas. value
The measurement is canceled when the specified measured value has
been reached since the start of the measurement.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the .
Measuring mode Ipol:
Input range –2000.0 - 2000.0 mV
Selection off
Default value off
Measuring mode Upol:
Input range
–200.00 - 200.00 µA
Selection off
Default value off
Measuring mode T:
Input range
–20.0 - 150.0 °C
Selection off
Default value off

Table of Contents

Other manuals for Metrohm 917

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 917 and is the answer not in the manual?

Metrohm 917 Specifications

General IconGeneral
BrandMetrohm
Model917
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals