28 Parameter
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
269
28.7.3.1 Dosing parameters
The parameters for the dosing are defined under [Dosing parameters].
Volume
Volume which is dosed.
Input range 0.00000 - 99999.9 mL
Default value 10.0000 mL
Dosing rate
The rate at which dosing takes place.
Input range 0.01 - 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum dosing rate depends on the cylinder volume (see chap-
ter 31, page 294).
Filling rate
Rate at which the dosing cylinder is filled.
Input range 0.01 - 166.00 mL/min
Selection maximum
Default value maximum
maximum
The maximum filling rate depends on the cylinder volume (see chapter
31, page 294).
28.7.3.2 Dosing device
The parameters for the dosing device are edited under [Dosing device].
Dosing device
Selection of the MSB connector to which the dosing device is connected.
The selection does not depend on the number of MSB connectors of the
control instrument.
Selection 1 | 2 | 3 | 4
Default value 1
Titrant
Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, etc.) is always used
for the calculation. Titrants are defined under System â–¶ Titrants. You
can also enter a name which is not contained in the titrant list. When a