28 Parameter
917 Coulometer
ï¿ï¿ï¿ï¿ï¿ï¿ï¿ï¿
275
Selection maximum
Default value maximum
maximum
The maximum dosing and filling rates depend on the cylinder volume
(see chapter 31, page 294).
28.7.4.2 Dosing device
The parameters for the dosing device are edited under [Dosing device].
Dosing device
Selection of the MSB connector to which the dosing device is connected.
The selection does not depend on the number of MSB connectors of the
control instrument.
Selection 1 | 2 | 3 | 4
Default value 1
Titrant
Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. This ensures that the correct data (titer, etc.) is always used
for the calculation. Titrants are defined under System â–¶ Titrants. You
can also enter a name which is not contained in the titrant list. When a
determination is started there is a check whether the solution is contained
in the list.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without an integrated
data chip, the cylinder volume and the type of dosing drive are checked.
For the selected titrant, the validity of the titer, the working life of the
titrant and the time interval for the GLP test of the buret unit are checked
at the start of the determination.
Selection
Selection of configured titrants | not defined
Default value not defined
not defined
No check takes place.