Monitoring Noninvasive Blood Pressure (NIBP) (Optional) Pairing a TM80 with a BP10
TM80 Telemetry Monitor Operator’s Manual 10 - 3
T.
10.2.2 System Responses after Successful Pairing
Once a TM80 is paired successfully with a BP10, the following system responses occur:
The icon will be changed to in the top-right corner of the TM80’s screen.
Besides, the appears in the top-right corner of the BP10.
The status of the BP10 which is paired successfully with the TM80 is displayed as
Connected. Besides, the “Pairing Successful” displays in the BP10’s message area
for about 10 seconds.
When NIBP measurement mode is set to ABPM at BP10 and the BP10 is paired
successfully with the TM80, the wireless seup confirmation message “Select Yes to
continue Patient in BP10. Select No to continue Patient in Telemetry.” will display
on the main screen of the TM80.
Ye s : Patient-related settings comes from BP10 and ongoing NIBP
measurement for current patient will not stop.
No: Patient-related settings at BP10 comes from the TM80. Besides, NIBP
measurement for the patient monitored by BP10 will stop and the data his-
tory related to this patient will be cleared at BP10. When a BP10 in ABPM
mode is not unpaired with a TM80 and needs to be paired with a new TM80,
you must select No. Otherwise, chaotic patient data may occur.
WARNING
Before pairing a BP10 with a new TM80, unpair it with the current TM80
first.
Do not pair a TM80 with an undesired BP10. If a TM80 is paired with an
undesired BP10, the patient category may be not applicable for the
patient and the measured NIBP data may be incorrect.
NOTE
Make sure that you have selected the correct BP10 to pair it with the
TM80.
If you pair a TM80 with a BP10 when NIBP measurement mode is set to
ABPM at BP10, the system responses after successful pairing are slightly
different from that when the NIBP measurement mode is set to Manual,
Auto, or Sequence. For the differences, see "System Responses after Suc-
cessful Unpairing" on page 10 - 5.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals