Technical Alarm Messages at the TM80 Troubleshooting
15 - 8 TM80 Telemetry Monitor Operator’s Manual
15.3 Technical Alarm Messages at the TM80
The following table lists the major technical alarm messages displayed in the prompt
message area of the TM80.
The Alarm Indication column in the table below is capable of three different indication
types: A, B, C.
ƾ A: The alarm sound will be silenced, the alarm light will continue to indicate the
alarm, a Ĝ will appear before the alarm message. The symbol appears on the
top of the main screen.
ƾ B: The technical alarm will be changed to the prompt message, it will not longer
make sound or be indicated by the alarm light.
ƾ C: The technical alarms are cleared, there will be no alarm indications.
NIBP NIBP-S
Extremely High
High The NIBP systolic
pressure has risen above
the extreme alarm high
limit or fallen below the
extreme alarm low limit.
Check the patient’s con-
dition and check if the
alarm limit settings are
correct.
NIBP-S
Extremely Low
NIBP-M
Extremely High
The NIBP mean pressure
has risen above the
extreme alarm high limit
or fallen below the
extreme alarm low limit.
NIBP-M
Extremely Low
NIBP-D
Extremely High
The NIBP diastolic
pressure has risen above
the extreme alarm high
limit or fallen below the
extreme alarm low limit.
NIBP-D
Extremely Low
Measurement Alarm message
Default
Alarm
priority
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.