Configuring the Alarms Menu Configuring the TM80
12 - 4 TM80 Telemetry Monitor Operator’s Manual
8. Select the desired option.
â—† Fixed: you cannot modify Room No. and Bed No. in the Patient Info menu.
Room No. and Bed No. are the ones entered in the Room No. and Bed No.
fields of the Device Location menu.
â—† Unfixed: you can modify Room No. and Bed No. in the Patient Info menu.
12.4 Configuring the Alarms Menu
1. In the Maintenance menu, tap Alarms.
The current setting displays to the right of the option.
2. Select the options described in the following table.
Section & Options Description Settings*
Sounds
Style Allows an authorized user to set the alarm tone
pattern.
ISO, Mode 1, Mode 2
Minimum Alarm Vol-
ume
Allows an authorized user to configure the
minimum alarm volume.
The minimum alarm volume defines the mini-
mum value you can set for the alarm volume.
For example:
If the minimum alarm volume is set to 5, the
minimum value you can set for the alarm vol-
ume in the Audio Volume menu is 5 (as shown
in the following figure).
If the minimum alarm volume is set to Off and
the Sounds is set to 0, the alarm sound is
turned off and the symbol appears on the
screen.
Off, 1, 2, 3, 4, 5, 6, 7, 8, 9, 10
Timeout
Reminder Tone Allows an authorized user to enable or disable
the reminder tone.
If the alarm sound is turned off, enabling this
setting can issue a periodic reminder tone.
On, Off
Reminder Interval Allows an authorized user to configure the
intervals between the alarm tones.
1 min, 2 min, 3 min, 5 min, 10 min
* The factory default settings are in bold.