Storing the Batteries Battery
14 - 6 TM80 Telemetry Monitor Operator’s Manual
Use the central charger to charge the lithium-ion batteries. The central charger can
charge 10 lithium-ion batteries at one time. For details about the central charger, refer to
BeneVision Central Charger Operator’s Manual (P/N: 046-007059-00).
14.7 Storing the Batteries
14.7.1 Storing Rechargeable Lithium-ion Batteries
When storing batteries, make sure that the battery terminals do not come into contact
with metallic objects.
If you need to store the batteries for an extended period of time, place the batteries in a
cool, dry place (ideally at 15°C or 60°F) with a partial charge of about 50% capacity (two
LEDs illuminated). Storing batteries in a cool place can slow the aging process.
Stored batteries should be charged to about 50% of their capacity every six months. The
battery should be fully charged prior to first use.
NOTE
• Remove the lithium-ion battery from the device if the device is not used
for a prolonged time (for example, several weeks), and keep the device
in clean place to avoid the dust or liquid entering the battery compart-
ment.
• Storing batteries at high temperatures for an extended period of time
will significantly shorten their life expectancy.
• Do not store the batteries in an environment above 60°C (122°F) or lower
than -20°C (4°F).

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals