Monitoring ECG, Arrhythmia, ST and QT ST Segment Monitoring
TM80 Telemetry Monitor Operator’s Manual 7 - 29
7.7.4.3 Arrhythmia Alarm Refractory Period
For the following eight alarms in medium priority chain, there is a 30-second refractory
period. That is to say, if the same alarm is not presented within 30 seconds after the pre-
vious alarm disappears.
ƾ Couplet
ƾ R on T
ƾ PVCs/min High
ƾ Multiform PVCs
ƾ PVC
ƾ Missed Beats
ƾ Pacer Not Capture
ƾ Pacer Not Pacing
7.7.5 Understanding the Arrhythmia Display
1. Parameter label
2. PVCs per minute label
3. PVCs per minute value
4. Pauses per minute label
5. Pauses per minute value
7.8 ST Segment Monitoring
ST segment analysis is intended for adult and pediatric patients.
When the TM80 is connected to the CMS, you can adjust ST points, enter the ST graphic
window and the ST view at the CMS. For more information, refer to "ST Monitoring" on
page 13 - 6.