Monitoring ECG, Arrhythmia, ST and QT ST Segment Monitoring
TM80 Telemetry Monitor Operator’s Manual 7 - 31
This example shows the ST digital area when the 6-lead ECG leadwires are used. Your
screen may look slightly different:
1. Parameter label
2. ST alarm off symbol
3. Lead label
4. ST unit
5. ST numerics: a positive value indicates ST segment elevation, and a negative value
indicates ST segment depression.
7.8.4 Configuring ST Alarm Settings
To configure ST alarm settings, follow this procedure:
1. Tap the ST digital area to enter the ST menu.
2. Tap ST Alarm Setup.
3. In the Setup section, tap ST Alarm Mode and then select the desired mode:
◆
Absolute: you can separately set the alarm properties for each ST alarm.
After selecting this option, the
Auto Limits option is displayed where
you can initiate auto adjustment of alarm limits.
◆ Relative: it is the default option. You can set the alarm properties for ST Sin-
gle and ST Dual alarms.
4. In the Alarms section, select the options described in the following table.