EasyManuals Logo

Mindray BeneVision TM80 User Manual

Mindray BeneVision TM80
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #229 background imageLoading...
Page #229 background image
Product Specifications Power Supply Specifications
TM80 Telemetry Monitor Operator’s Manual A - 3
A.3 Power Supply Specifications
A.3.1 TM80
The battery life listed below may be shortened by the following factors:
â–  Wireless signal interference
â–  Aging battery
â–  Frequent operation of turning on/off the display
â–  Wireless network coverage which does not meet the technical specifications in
"Wireless Specification" on page A - 6.
Battery type
AA batteries (three)
Rechargeable lithium-ion battery (one)
Capacity for lithium-ion
battery
≥ 3500 mAh
Charge time for lithium-ion
battery
At the room temperature: ≤ 5 hours to 90% charge
Safety and authentication
for lithium-ion battery
UL2054, IEC62133, UN38.3
Battery run time
(at 25 °C ±5 °C, with display
off, no sound and light as
default, no alarms, display
lit for one minute every two
hours, using L91 AA batter-
ies (PN 0146-00-0077-10)
Three AA batteries Lithium-ion battery
5-lead ECG ≥36 hours ≥36 hours
5-lead ECG+SpO
2
≥28 hours ≥28 hours
Shutdown delay
At least 10 minutes after the low battery alarm first occurs.
At least five minutes after the critically low battery alarm first occurs.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals