Monitoring Respiration (Resp) (Optional) Understanding the Resp Display
TM80 Telemetry Monitor Operator’s Manual 8 - 7
8.5 Understanding the Resp Display
8.5.1 Resp Digital Area
1. Parameter name
2. Measurement unit
3. Respiration rate (RR)
4. Alarm high limit and low limit.When RR alarm is switched off, the alarm off symbol
instead of alarm limits is displayed.
8.5.2 About the Resp Digital Area
■ The Resp digital area displays the respiration rate in the unit of RPM with a
resolution of 1 RPM.
■ If the Resp measurement is invalid, “- - -” displays in place of the RR value.
No Chest Movement The TM80 will trigger the “No Chest Movement”
alarm if the patient has stopped breathing for
longer than the set apnea time. This option
defaults to be switched on.
When the TM80 is connected to the Central
Monitoring System, switching on or off this
option at the Central Monitoring System also
switches it on or off at the TM80, and vice versa.
After tapping this option, you can configure the
alarm priority.
Alarm Priority: High
Options Description Settings*
* The factory default settings are in bold.