Pausing Alarms Alarms
6 - 8 TM80 Telemetry Monitor Operator’s Manual
6.5.4 Restoring Default Alarm Settings
To restore default alarm settings, follow this procedure:
1. In the main menu, tap Alarms.
2. Tap Alarm Limits.
3. In the Setup section, tap Defaults.
4. In the dialog that pops up, tap OK.
6.6 Pausing Alarms
When an alarm occurs, follow this procedure to pause the alarm system.
1. Press the key to enter the main menu.
2. In the Commands section, tap Alarm Pause.
OR
1. Swipe your finger up at the bottom of the main screen to display the quick keys
area.
2. Tap Alarm Pause.
When alarms are paused, the following indicators are given:
No physiological alarm will be presented within the alarm pause time.
For technical alarms, alarm sounds are paused, but alarm lamps and alarm
messages continue to be presented.
The remaining alarm pause time is displayed in the prompt message area.
The alarm pause symbol is displayed in the upper right corner of the main
screen.
The alarm pause time is two minutes by default. You can change the alarm pause time at
the CMS. For details, refer to BeneVision Central Monitoring System Operator’s Manual.
When the pause time expires, the alarm paused status is automatically deactivated.
NOTE
When the TM80 is connected to the CMS and the function of remotely
pausing alarms is enabled at the CMS, alarms can be paused either at the
TM80 or at the CMS. For information on pausing alarms at the CMS, refer
to BeneVision Central Monitoring System Operator’s Manual.