Pairing Introduction Monitoring Noninvasive Blood Pressure (NIBP) (Optional)
10 - 2 TM80 Telemetry Monitor Operator’s Manual
10.1 Pairing Introduction
NIBP (non-invasive blood pressure) data is measured by BP10 using the oscillometric
method. This measurement can be used for adult and pediatric patients.
You can connect a TM80 to a BP10 via Mindray Patient Area Network (abbreviated as
MPAN). The process of establishing connection between the TM80 and the BP10 is called
“pairing”.
When the TM80 is paired with the BP10, measured NIBP data and NIBP-related technical
alarms will be transferred from the BP10 to the TM80. You can view NIBP data and NIBP-
related technical alarms and physiological alarms on the TM80’s screen.
For MPAN specifications, refer to "Measurement Specifications" on page A - 10. For the
functionalities and operations of the BP10, refer to BP10 NIBP Module Operator’s Man-
ual.
10.2 Pairing a TM80 with a BP10
10.2.1 Pairing Procedures
Before pairing the TM80 with a new BP10, first unpair the already connected BP10 by fol-
lowing steps in "Unpairing the TM80 with the BP10" on page 10 - 4.
To pair a TM80 with a BP10, follow this procedure:
1. Press the MPAN key on the right panel of the BP10. The “Pairing...” message will dis-
play in the prompt message area of the BP10.
2. Press the key to enter the main menu of the TM80.
3. Tap Wireless Modules.
4. Tap MPAN, and then select Connect to BP10.
The TM80 starts searching the BP10 devices in the vicinity and displays the devices
that can be connected under Devices. Make sure that the device you wish to con-
nect appears in the list of devices. If not, repeat Steps 1 to 4.
5. In the list of devices, select your desired device and tap Connect. Once the device
is connected successfully, the status will be changed from Disconnect to Con-
nected.
6. Tap anywhere in the NIBP parameter area on the TM80’s main screen to access the
NIBP menu.
7. Tap Start to start NIBP measurement. Verify that the NIBP measurement results
displayed at the BP10 are consistent with that displayed on the main screen of the
TM80 and are the data for the same patient.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals